Water
is utilized abundantly to process, formulate, and manufacture
pharmaceutical products. Traditional culture-based methods used to
ensure water quality, however, are ill-suited in providing a robust
assessment of risk and control. These methods are plagued by limitations
in sensitivity, episodic sampling, and retrospective results. New
technologies based on laser-induced fluorescence (LIF) detect intrinsic
fluorescence instead of growth, can operate continuously, and deliver
real-time results. As applied to pharmaceutical water quality,
LIF-based, instantaneous microbial detection technologies enable
real-time bioburden monitoring, risk reduction, and process control.
Figure
1: Emission spectra of two microorganisms and eight materials with
405nm excitation. An approximate Raman band for water and two example
detection ranges for an instantaneous microbial detection system with
two PMTs are labeled.The pharmaceutical industry
continues to recognize a need to leverage modern technologies to advance
the course of risk reduction and process control. This forward thinking
has been captured in industry relevant guidance such as the FDA’s 2004
“Guidance for Industry” document on Process Analytical Technology (PAT),
ICH Guidelines Q8, Q9, and Q10, and the FDA’s “Pharmaceutical cGMPs for
the 21st Century,” which encourage the adoption of quality by design
(QbD) principles and new technologies. More recently, working groups
composed of representatives from key pharmaceutical companies have also
joined forces to help articulate their needs in water quality assessment
and encourage the development and use of new technologies best suited
to today’s tasks.
Need for an online pharmaceutical water assessment tool
The
currently accepted and primarily practiced method for assessing water
quality throughout a pharmaceutical water loop is through samples
obtained at points-of-use (POU), utilizing traditional culture-based
methods. The goal of such testing is to ensure the quality of an entire
water system; however, POU testing can occur as infrequently as once
every two weeks at each sample point. This limited sampling frequency,
combined with the retrospective nature of culture-based methods, make a
robust and timely assessment of risk and control difficult. Furthermore,
there is the potential for sample contamination during collection (a
false positive), and for a false-negative result due to limitations in
sensitivity of culture-based methods. While growth-based methods offer
the opportunity for identification, a number of organisms go undetected,
such as viable but non-culturable organisms, due to the chosen medium
and incubation parameters. A complementary technology capable of
real-time and continuous monitoring of water system bioburden, based on a
different method of detection, could alleviate such limitations and aid
in risk reduction and process control.
Table 1: Business benefits summarized in the OWBA Business Benefit Estimates document.Online pharmaceutical water bioburden analyzer
With
an aim to improve the tools being applied to pharmaceutical waters, an
Online Water Bioburden Analyzer (OWBA) Workgroup recently outlined user
requirements, a testing protocol, and business benefits to guide the
development of an OWBA system.1,2,3,4 This workgroup,
composed of representatives from seven major pharmaceutical companies,
has a mission to aid instrumentation vendors in the creation of an
online water bioburden analyzer that satisfies both industry and
regulators. They believe, “an online water bioburden analyzer has the
potential to eliminate sampling and testing errors via reduced
manipulations while providing increased product safety and process
control through the availability of statistically significant data.”3
According to the group, such an OWBA system is not primarily designed
to eliminate compendial water testing, but should be used as a risk
reduction tool. Potential business benefits are shown in Table 1 and
include energy savings, labor reduction (resource allocations), and
increased product quality and process understanding.2
Figure
2: With a scatter detector and two fluorescence detectors (PMTs), an
instantaneous microbial detection system for water can create a
three-dimensional plot of biologic and interferent particles. Through
assessment of the three different signals and an advanced processing
algorithm, such a system offers enhanced interferent discrimination
capabilities.Technical system requirements are
provided, which include specifications for bioburden sensitivity,
calibration, chemical compatibility, operating parameters, and needed
consumables.3 Also included is a requirement for a limit of detection
(LOD) equivalent to that set forth for culture-based methods (10
CFU/100mL) and analysis modes that include continuous sampling,
time-based sampling, and daily operation at designated times. Overall,
the system should be capable of continuous and periodic monitoring of
critical control points (CCP) and POU, with sufficient sensitivity to
detect microorganisms in water and limited susceptibility to potential
interferents such as rouge, residual sanitizer, and gasket materials.
Laser-induced fluorescence
One
technique capable of satisfying the OWBA requirements is laser, or
light, induced fluorescence (LIF). LIF is a spectroscopic technique
capable of high sensitivity in the detection of compounds that
fluoresce. Fluorescence is the luminescence that occurs with the
absorption of radiation at one wavelength followed by the emission of
radiation at a different wavelength. Substances that typically fluoresce
may be referred to as fluorophores. Quinine is a familiar fluorophore
due to its presence in tonic water.
The application of LIF to
detect microorganisms has been leveraged in flow cytometry, capillary
electrophoresis, solid-phase cytometry, adenosine triphosphate
bioluminescence, and growth-based auto fluorescence. In a number of
these techniques, microorganisms are dyed to increase the measurable
fluorescence. Measuring the intrinsic fluorescence of a microorganism
removes the requirement for dyes and sample preparation, but requires an
instrument with significant sensitivity. As lasers of additional
wavelengths at higher power levels have become commercially available,
LIF has become very relevant in applications requiring detection of low
levels of microbial intrinsic fluorescence.
A light source such
as a laser is the excitation source in LIF. A laser of appropriate
wavelength and intensity is capable of inducing intrinsic fluorescence
emission from microbes due to constituent fluorophores such as
tryptophan, nicotinamide adenine dinucleotides (NADH), and flavins that
are present in microorganisms.7 The target excitation
wavelength is based on the excitation spectra of target fluorophores
such that sufficient fluorescence intensity is induced for measurement
and a greater number of non-biologic materials may be excluded. Yet,
non-biologic materials such as plastics, rubbers, and paper can also
fluoresce pointing to the importance of software discrimination
algorithms.
OWBA: Instantaneous microbial detection technology for water
An
OWBA system based on LIF enables the instantaneous detection of
microbes in water, without the need for consumables and the limitations
presented by traditional testing methods. Commercially available systems
for water employ a 405nm laser to simultaneously induce Mie scatter and
intrinsic fluorescence, on a particle-by-particle basis, as a sample
travels along a flow path and traverses this excitation source.
Detection and correlation of the Mie Scatter and fluorescence signals
provide real-time information on the presence and biologic status of
particles. Detection based on the intrinsic fluorescence of
microorganisms removes requirements for sample preparation. Furthermore,
this fundamental method of detection is inherently different from
traditional growth-based methods, and is not susceptible to the
growth-based limitations resulting from improper media selection and
incubation.
Figure
3: Representative data from the IMD-W system showing IMD-W biologic
counts as compared to colony forming unit culture results obtained using
the traditional method with TSA plates.In
LIF-based systems, intrinsic fluorescence is captured on a
photomultiplier tube (PMT), a detector highly sensitive to light. Both
one-PMT and two-PMT designs are available. In water, two-PMT designs
provide better discrimination of non-biologic fluorescing materials such
as rouge, as requested in the OWBA requirements.3 Each
material and microorganism has a different excitation and emission
spectrum. Once an excitation wavelength has been chosen, some materials
show a broad fluorescence emission and others a narrow emission
spectrum. Similarities in the emission spectra of biologic versus
non-biologic materials can be used advantageously. Figure 1 contains the
emission spectra of certain biologic and non-biologic materials with
405nm excitation. Two notional PMT detection regions have been
highlighted on either side of a Raman band in this figure. The Raman
band represents fluorescence produced from the interaction of the laser
light with water. Therefore, in order to detect particulate within the
interrogated water stream, this band must be avoided in the detection
regions utilized by the system. With 405nm excitation, the Raman band
for water has a maximum at approximately 469nm8.
With
two PMT detection regions, the differences in non-biologic versus
biologic emission spectra can be utilized to aid in the classification
of non-biologic materials as inert. As shown in Figure 2, a particle’s
scatter and fluorescence signals can be combined to create a
three-dimensional map of interferent and biologic particles. Advanced
algorithms can then be utilized to aid in the discrimination of biologic
and interferent materials.
Real-time bioburden monitoring, risk reduction, and process control
The
use of an instantaneous microbial detection system for pharmaceutical
water provides the ability to monitor bioburden continuously and in real
time, resulting in an increased potential for risk reduction and
process control. Figure 3 shows representative data from the IMD-W™, a
system designed with the OWBA requirements in mind, comparing IMD-W
biologic counts to culture plate results for three OWBA suggested
organisms. This data covers a wide dynamic range and speaks to the
potential sensitivity and ability of such systems to monitor bioburden.
The
continuous data offered by these systems creates a robust historical
dataset that is ideally suited for trending, particularly when compared
to episodic sampling with traditional methods. Sampling considerations
set forth in “USP<1231> Water for Pharmaceutical Purposes”
recommends monitoring pharmaceutical water systems at a frequency
“sufficient to ensure that the system is in control and continues to
produce water of acceptable quality.”5 The general
information chapter states it is best to operate monitoring
instrumentation in a continuous mode such that a large volume of
in-process data can be generated, and suggests the use of trend analysis
as an alert mechanism for loop maintenance.5 A combination
of historical trending data and real-time results enable users to
identify an out-of-specification event or deterioration in
microbiological control significantly earlier than with traditional
sampling methods. By continuously monitoring the state of control,
timely loop maintenance can be performed if bioburden data trends
upward, permitting further risk reduction and an increased level of loop
control. A real-time and historical knowledge of control can also be
important during a POU testing deviation.2 If POU testing is
positive for microbial contamination, knowledge and data to support a
state of control may narrow the root-cause investigation to the POU as
opposed to contamination in the entire loop.
Conclusions
Regulatory
guidance and calls from industry work groups support the need for
better tools for pharmaceutical water monitoring. New instantaneous
microbial detection systems based on LIF enable real-time bioburden
monitoring, increased risk reduction, and process control for
pharmaceutical waters. Through continuous monitoring, these systems
provide significant historical data for robust trending and assessment
of water loop bioburden levels, providing the means to monitor the level
of control and react to out-of-specification events in a much more
timely manner than with traditional methods alone. Users stand to
benefit through increased product quality and process understanding,
energy savings, and risk reduction.
References 1. Cundell, A., Luebke, M., Gordon, O., Mateffy, J., Haycocks, N., Weber, J. W., et al. (2013, May 16). On-Line Water Bioburden Analyzer Business Benefits Estimation.
Retrieved August 8, 2014, from
http://www.miclev.se/fileadmin/user_upload/jennie/Online_Water_BioBurden_Analyzer_Business_Benefits.pdf
. 2. Cundell, A., Gordon, O., Haycocks, N., Johnston, J., Luebke, M., Lewis, N., et al. (2013, May/June). Novel
Concept for Online Water Bioburden Analysis: Key Considerations,
Applications, and Business Benefits for Microbiological Risk Reduction.
American Pharmaceutical Review, 26-31. 3. Cundell, A., Luebke, M., Gordon, O., Mateffy, J., Haycocks, N., Weber, J. W., et al. (2013, March 18). On-Line Water Bioburden Analyzer User Requirement Specifications (URS). Document ID OWBA-DURS-2013-v1.3. 4. Cundell, A., Luebke, M., Gordon, O., Mateffy, J., Haycocks, N., Weber, J. W., et al. (2013, April 24). On-Line Water Bioburden Analyzer Testing Protocol. Document ID OWBA-TP-2013-v1.5. 5.
USP<1231> Water for Pharmaceutical Purposes. Pharmacopeial forum,
Vol. 32; United States Pharmacopeial Convention, Inc.: Rockville, MD,
2008. 6. Lakowicz, J. R. (2006). Principles of Fluorescence Spectroscopy (3rd ed.). New York: Springer Science & Business Media. 7. Ammor, M. S. (2007). Recent Advances in the Use of Intrinsic Fluorescence for Bacterial Identification and Characterization. Journal of Fluorescence, 17:455-459. 8. Rouessac, F., & Rouessac, A. (2013). Chemical Analysis: Modern Instrumentation Methods and Techniques (2nd ed.). West Sussex: John Wiley & Sons.
\\1231>1231>
No comments:
Post a Comment