Refine testing to meet FDA concerns about over-the-counter formulations
In September, the FDA and U.S. Pharmacopeial Convention held a joint
meeting to discuss limitations of the testing procedures for
over-the-counter drug substances and products. Reviews of testing
protocols for OTC products found that many are not adequate for testing
for such things as impurities, which would indicate expired products.
Developing suitable testing methods can be more challenging than it
might initially seem. Many OTC cough and cold medicines, for example,
contain more than one API, which, due to differences in efficacy, are
dosed at different concentrations. Thus, developing a single test for
all degradants in one analysis can be challenging.
High performance liquid chromatography—more specifically,
reversed-phase HPLC—is the technique of choice for quantitation of drug
products and impurities. Other techniques, such as ion exchange, do not
provide the efficiency necessary to quantitate impurities at the
required 0.1% level. Fortunately, there are many different RP-HPLC
stationary phases and particle types available today.
When dealing with formulations containing multiple active components,
it is often beneficial to utilize high-efficiency particles that will
give improved resolution between critical components. Kinetex
(Phenomenex) core-shell particles were engineered to provide increased
efficiencies and improved performance compared to traditional fully
porous particles. This improved performance typically leads to much
faster separations and higher sensitivity.
click for larger view
FIGURE 1. Analysis of
Cough and Cold Medicine (NyQuil) on Kinetex 2.6μμm C18 100 x 4.6mm ID.
The high-efficiency nature of the Kinetex particle allows for the
accurate quantitation of all drug substances in less than four minutes.
Comparing results from complementary stationary phases such as the
C18, XB-C18, and PFP can help identify the best separation. In the case
of the well-known cough and cold formulation, Vicks NyQuil, the Kinetex
C18 provided the best resolution and fastest analysis time for all three
active components (Figure 1).
Difficult Formulations
Adding more complications are the various formulations for the same
medicine: tablet, gel capsule, liquid syrup, and suspensions. In any of
these dosage forms, the API or APIs usually only make up a small
fraction of the total mass, with the bulk being made up of various
excipients such as stabilizers, fillers, flavoring agents, and
preservatives. It is the fillers and excipients that are often the
source of many of the chromatographic challenges.
Oral drug suspensions can be especially problematic for HPLC analysis
due to the fact that, by definition, a suspension is a fluid that
contains undissolved solids, which are large enough to undergo
sedimentation. When injected onto an HPLC, the undissolved solids can
physically clog the inlet frits of the column, the pores of the silica,
and the interstitial space between the silica particles. Physical
clogging of the media and/or frits most often manifests as an increase
in system backpressure, which, if it gets high enough, can shut down an
HPLC run and also render the HPLC column unusable.
click for larger view
FIGURE 2. A comparison of
the Effectiveness of Centrifugation Versus a Two-Step Filtration for A
Drug Suspension Sample. Filtering a drug suspension sample through a
0.45 µm followed by a 0.2 µm PTFE filter is much more effective at
protecting the analytical HPLC column from clogging due to suspended
particulates in the suspension matrix than is centrifugation alone.
Depending upon the nature of these undissolved solids, they may be
readily removed using centrifugation. Centrifugation of a diluted
aliquot of a drug suspension, using a small benchtop centrifuge for five
to 10 minutes, can remove suspended particulates. However, Figure 2
gives an example of a centrifuged OTC suspension where injection of the
supernatant led to an increase in system backpressure with each
sequential injection. Upon removal of the HPLC column and replacement
with a new column, the system pressure returned to normal, indicating
that the pressure buildup was within the column itself and not in the
HPLC system plumbing. Back-flushing of the column using a highly aqueous
solution and a 100% organic solvent (acetonitrile) for about one hour
resulted in only a partial return to the initial pressure, indicating
that, for all practical purposes, the clogging was irreversible.
The use of syringe filters for removing particulates from samples
prior to injection is well established, and there are many materials to
choose from. In the case of the previous OTC suspensions, we have found
that filtration using a Phenex 0.45 µm polytetrafluoroethylene filter
followed by a second filtration using a Phenex 0.20 µm PTFE filter was a
more effective way to remove the undissolved microparticulates. After
this process, even relatively large sample volumes (50-100 µL if
necessary) can be injected without significantly affecting system
backpressure.
Filtering with a Phenex 0.45 µm filter alone will not usually protect
the column from clogging. Likewise, skipping the first crude filtration
step and attempting to filter the initial drug suspension (or even a
10-fold dilution of drug suspension) directly through a Phenex 0.2 µm
filter will typically result in the filter itself rapidly becoming
clogged before a sufficient volume of the drug suspension has been
clarified.
click for larger view
FIGURE 3. Effectiveness
of Security Guard Ultra at Protecting an Analytical Column from Physical
and/or Chemical Degradation. In this accelerated longevity test using
sequential injections of an unfiltered matrix, the guard cartridge
system is able to effectively absorb the majority of the contaminants
that would otherwise foul the analytical column.
Filters are designed to be inert, but they can interact with the
drug, causing problems with recovery. If a filter material is to be used
for an assay, a filter absorption study must be done to verify recovery
of the API or its associated impurities. If recovery is too low, it may
be necessary to explore alternative membrane types such as nylon.
Please note that, while filtration and centrifugation are effective
at removing undissolved solids from drug suspensions, neither is
effective for removing the surfactants and emulsifiers that are often an
integral part of the suspension formulation. Thus, while filtration
will often protect that HPLC column from exhibiting increases in
backpressure due to clogging, it is still quite common to see a gradual
shift in chromatography (retention, selectivity, and so on) that can be
attributed to the gradual buildup of surfactants from the sample matrix
in the HPLC column. An additional sample preparation step is necessary
to deal with these intractable components of the suspension matrix.
Surfactants and Emulsifiers
In order to maintain the relatively homogenous composition of the
drug suspension and prevent, as much as possible, the components in the
formulation from settling out, drug manufacturers must add a variety of
surfactants and/or emulsifiers. Surfactants, which display a high
affinity for the C18 phases that are typical in RP-HPLC methods, can be
very difficult to wash off under normal mobile phase conditions and tend
to build up over time, gradually changing the nature of the stationary
phase surface and leading to a shift in chromatographic retention and
selectivity. This gradual shift in chromatography as a result of
surfactants from the sample matrix is problematic in that it may cause
an HPLC column to fail system suitability testing if a monitored
parameter, such as retention time or resolution, shifts outside of an
acceptable limit.
Ionic surfactants can be removed using ion-exchange SPE resins that
selectively retain the surfactant, allowing the API and related
substances to pass through. However, many non-ionic surfactants such as
Tween (polysorbate) cannot be removed by this or any other mechanism. In
such cases, the use of a guard column can be highly effective in
preserving column performance.
click for larger view
FIGURE 4. A Guard Column
Can Protect the Analytical HPLC Column from Shifts in Chromatography Due
to Injection of Drug Suspension Samples that Contain Emulsifiers and/or
Surfactants. Injection of a drug suspension sample containing the
surfactant Tween results in an obvious and rapid shift in
chromatographic retention for several of the peaks (top overlay; red
trace = initial rejection, blue trace = second injection of drug
suspension containing Tween-80). However, with the C18 guard cartridge
system in place, the shift in retention is not apparent.
Guard columns protect analytical columns in two ways: 1) by
physically capturing suspended particulates that would otherwise clog
the analytical column and 2) by trapping chemicals from either the
mobile phase or the injected sample that may have a detrimental effect
on HPLC column performance (Figure 3). Cartridge-based systems such as
the SecurityGuard and SecurityGuard ULTRA are a cost-effective way to
protect the analytical column. When changed out at appropriate time
intervals, the same analytical column can be used for a longer time than
would be possible without the guard column.
When working with drug suspensions, guard columns are particularly
important. We have found guard columns to be vital in preserving column
performance. SecurityGuard C18 appears to absorb some of the emulsifier
and/or surfactants present in a sample matrix, preventing them from
accumulating in the analytical column and preserving chromatography
(Figure 4). In the absence of a SecurityGuard C18 guard, a single
injection of the placebo matrix triggers a substantial change in
chromatography, which continues to degrade performance on subsequent
injections. However, when the guard column is included, the injections
of placebo have little if any effect on chromatography.
We must attribute this protective effect to the guard column
“capturing” some surfactant contained in the injected sample matrix. Of
course, the ability of the guard to absorb these detrimental components
will be determined by the relative amount of media present in the guard
and the amount of contaminant loaded with each injection. To prevent
saturation of the guard column, we recommend changing it at regular
intervals, a practice facilitated by the use of disposable
cartridge-type guard column systems.
The FDA is heavily scrutinizing OTC formulations. The USP and other
organizations are working hard to develop standard testing methods for
companies to use. In the meantime, robust analytical methods for common
OTC formulations are needed.
We have found guard columns to be vital in
preserving column performance. SecurityGuard C18 appears to absorb some
of the emulsifier and/or surfactants present in a sample matrix.
OTC
drug formulations present an extreme challenge for the HPLC analyst due
to the presence of multiple active components and the complex nature of
the sample matrix, which make it difficult to separate active
ingredients and their impurities from excipients. The use of
high-efficiency Kinetex core-shell materials provides increased
resolution that can help resolve critical impurities.
When working with challenging formulations like suspensions, the
presence of particulates and emulsifiers in the sample matrix can lead
to poor column lifetime and reproducibility. When working with these
types of formulations, analytical HPLC column performance and lifetime
can be extended significantly with the use of Phenex syringe filters
(0.45 µm and 0.2 µm PTFE), which physically capture the suspended
particulates before they can clog up the analytical column. The use of
the SecurityGuard and SecurityGuard ULTRA guard cartridge systems (C18
cartridge-type preferred) can also chemically absorb surfactants and
emulsifiers that can alter HPLC column performance over time.
Jeff Layne, PhD, is technical manager for PhenoLogix at Phenomenex, and Sky Countryman manages PhenoLogix and applied technology for the firm.
Editor’s Choice
- Aicher B, Peil H, Peil B, Diener HC. Pain measurement: Visual
Analogue Scale (VAS) and Verbal Rating Scale (VRS) in clinical trials
with OTC analgesics in headache. Cephalalgia. 2012;32(3):185-197.
- Friedland S, Kothari S, Chen A, Park W, Banerjee S. Endoscopic
mucosal resection with an over-the-counter hyaluronate preparation
[published online ahead of print Feb. 29, 2012]. Gastrointest Endosc.
- U.S. Food and Drug Administration. Drug applications for over-the-counter drugs. FDA website. Available at: www.fda.gov/drugs/developmentapprovalprocess/howdrugsaredevelopedandapproved/approvalapplications/
over-the-counterdrugs/default.htm. Accessed March 13, 2012.
|
No comments:
Post a Comment